881/882 Video Test Generator User Guide (Rev A.22) 79
8. To save the changes, press the Enter (Options) key. The following choices appear on
the generator’s display:
• To save the changes, choose the Yes item by pressing the adjacent soft key.
• To exit without saving the changes, choose the No item.
• To return to the previous screen without saving the changes, choose the Back item.
Using the ImageShift utility through the command line interface
The procedures for configuring and running the ImageShift utility from the command line
interface are given below. To run the ImageShift utility from the command line interface,
you must utilize an imageshift file. An imageshift file specifies both the image and
parameters necessary to run the ImageShift utility.
To create an imageshift file:
1. Establish a session with the generator using either HyperTerminal over a serial
connection or Telnet over an Ethernet LAN. See “Configuring the generator’s serial
port” on page 32 or “Establishing a Telnet session with the generator” on page 34.
2. Set the path to the imageshift files using the ISHP command. For example:
ISHP tffs0/library/imageshifts
3. Create a new imageshift file using the ISHN command. For example:
ISHN Newshift
4. Define the imageshift file settings as shown in the example below:
ISHB //begins an imageshift editing session
XISH:SRCN /cache0/images/SMPTEBar //selects the SMPTEBar image
XISH:PATH 0,0 150,20 //sets the imageshift path x,y coordinates
XISH:TTYP 1 //sets the trace type to Repeat (1)
XISH:HINC 4 //sets the horizontal shift increment to 4
XISH:VINC 2 //sets the vertical shift increment to 2
XISH:TINC 5 //sets the time shift value to 5
ISHE //ends the imageshift editing session
5. Save the settings to the same imageshift file using the ISHS command, For example:
ISHS //saves the settings to the Newshift file
Apply Settings?
Back
Yes No