Quantum-data 802BT Bedienungsanleitung Seite 109

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 744
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 108
802BT/802R Video Test Generator User Guide 99
Setting the GPIB port address
The default GPIB address is 15. You can specify a different address.
To set the address of the GPIB port directly on the generator:
1. Locate the rotary switches on the back of the generator between the RS-232 port and
the GPIB port. These are accessible through the ventilation slots. There are two rotary
switches: the one on the left configures the value of the tens digit and the one on the
right configures the value of the ones digit.
2. Set the rotary switches first the value of the tens switch and then the value of the ones
switch to the desired value.
3. Hold down the DCS key, and then cycle the power to the generator. The port is set to
the new address.
Note: A normal power cycle will not update the memory to the new switch settings.
To set the address of the GPIB port using the command line:
1. Establish a terminal session with the generator
2. (Optional) Enter the following command to check the current address:
GPIB?
3. Enter the following command:
GPIB address
When the address is changed with the GPIB command, the change takes place as
soon at the command is issued.
The new address setting will be maintained until 1) the address is changed with another
GPIB command, 2) the unit is re-initialized with an INIT command, or 3) the unit is
re-initialized because the power-up self test found corrupted data in battery-backed
system memory. In cases 2 and 3, the GPIB address will revert to the address set on
the GPIB address switches if the unit has them, or to 15 if the unit does not have GPIB
address switches.
Seitenansicht 108
1 2 ... 104 105 106 107 108 109 110 111 112 113 114 ... 743 744

Kommentare zu diesen Handbüchern

Keine Kommentare